Validating a computer aided glaucoma screening system using RIGA 2 dataset

Medical Imaging 2021: Imaging Informatics for Healthcare, Research, and Applications(2021)

引用 0|浏览1
暂无评分
摘要
The purpose of this study was to evaluate the ability of a developed computer aided glaucoma screening system to screen for glaucoma using a Food Drug Administration (FDA) Class II diagnostic digital fundus photography system used for diabetic retinopathy screening (DRS). The fundus photos collected from participants underwent a comprehensive eye examination as well as non-mydriatic 45 degrees single photograph retinal imaging centered on the macula. Optic nerve images within the 45 degrees non-mydriatic and non-stereo DRS images (The Retinal fundus Images for Glaucoma Analysis: the RIGA2 dataset) were evaluated by a computer-aided automated segmentation system to determine the vertical cup-to-disc ratio (VCDR). The VCDR from clinical assessment was considered as gold standard, VCDR results from the computer system was compared to that from clinical assessment. The grading agreement was assessed by computing intraclass correlation coefficient (ICC). In addition, sensitivity and specificity were calculated. Among 245 fundus photos, 166 images met quality specifications for analysis. The ICC value for the VCDR between the gold standard clinical exam and the automated segmentation system was 0.41, indicating fair agreement. The specificity and sensitivity for (0.6 VCDR) were 76% and 47% respectively.
更多
查看译文
关键词
Telemedicine, tele-glaucoma, automated screening system, glaucoma screening
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要