Traceable dosimetry for MeV ion beams

G Garty, A D Harken,D J Brenner

JOURNAL OF INSTRUMENTATION(2022)

引用 1|浏览2
暂无评分
摘要
Standard dosimetry protocols exist for highly penetrating photon and particle beams used in the clinic and in research. However, these protocols cannot be directly applied to shallow penetration MeV-range ion beams. The Radiological Research Accelerator Facility has been using such beams for almost 50 years to irradiate cell monolayers, using self-developed dosimetry, based on tissue equivalent ionization chambers. To better align with the internationally accepted standards, we describe implementation of a commercial, NIST-traceable, air-filled ionization chamber for measurement of absorbed dose to water from low energy ions, using radiation quality correction factors calculated using TRS-398 recommendations. The reported dose does not depend on the ionization density in the range of 10-150 keV/mu m.
更多
查看译文
关键词
Detector alignment and calibration methods (lasers, sources, particle-beams), Dosimetry concepts and apparatus, Gaseous detectors, Very low-energy charged particle detectors