Pilot study for the comparison of machine-learning augmented audio-uroflowmetry with standard uroflowmetry in healthy men

BMJ Innovations(2020)

引用 6|浏览4
暂无评分
摘要
Background Routine assessments of lower urinary tract symptoms (LUTS) include standard uroflowmetry (UF), which is labour and equipment intensive to perform, and stressful and unnatural for patients. An ideal test should be accurate, repeatable, affordable and portable. Objective To evaluate the accuracy of a machine-learning (ML) augmented audio-uroflowmetry (AF) algorithm in predicting urinary flows. Subjects and methods This pilot study enrolled 25 healthy men without LUTS, who were asked to void into a gravimetric uroflowmeter. A smartphone recorded the voiding sounds simultaneously. Paired uroflow and audio parameters were used to train an ensemble ML model to predict urinary flows from voiding sounds. Pearson’s correlation coefficient was used to compare UF with AF values. Statistical significance was defined as p Results A total of 52 voiding session were captured, of which n=35 were used for training and n=17 for testing the algorithm. Each voiding session was divided into 0.1 s frames, resulting in >300 analysable datapoints per session. Pearson’s coefficients showed strong correlations for flowtimes (r=0.96, p Conclusions ML augmented AF can predict uroflow parameters with a good accuracy, and can be a viable alternative to standard UF. Further work is needed to develop this platform for use in real-life conditions and across genders.
更多
查看译文
关键词
standard audio-uroflowmetry,healthy augmented,machine-learning machine-learning
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper