Glassy carbon electrode modified with ex-in-situ gold film – A simple and effective working electrode for As(III) determination by using differential pulse anodic stripping voltammetry

Hue University Journal of Science: Natural Science(2022)

引用 1|浏览1
暂无评分
摘要
An easy-to-make new working electrode, an ex-in-situ AuF/GCE, was developed for trace As(III) detection. A gold film electrode prepared ex-situ was re-plated in-situ during each arsenic deposition step by adding Au(III) into the analyte solution. The factors affecting arsenic stripping response, namely, gold film preparation conditions, electrolyte concentration, electrode cleaning potential, cleaning time, deposition potential, and deposition time, were investigated. Compared with the traditional gold film electrodes prepared ex-situ, the new electrode has better precision and linearity of arsenic differential pulse anodic stripping voltammetry responses. For a deposition time of 90 s at –200 mV, the new electrode exhibits a sensitivity, a limit of detection (3-Sigma), a limit of quantitation of 0.103 μA·L·μg–1, 0.4 μg·L–1, and 1.3 μg·L–1, respectively.
更多
查看译文
关键词
arsenic,As(III),arsenic determination,anodic stripping voltammetry,gold film
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要