Quantitative analysis of vitamin D using m/MALDI-TOF mass spectrometry based on a parylene matrix chip

Journal of Analytical Science and Technology(2022)

引用 2|浏览1
暂无评分
摘要
Vitamin D deficiency is associated with various disorders and is diagnosed based on the concentration of 25-hydroxy vitamin D3 (25(OH)D 3 ) in serum. The parylene matrix chip was fabricated to reduce the matrix background noise, and the homogenous distribution of the matrix was retained for the quantitative analysis of 25(OH)D 3 . The Amplex Red assay was performed to confirm that the sample-matrix mixing zone of the parylene matrix chip was formed below the surface of the parylene-N film. The homogeneous distribution of the matrix was verified from the fluorescence image. For effective analysis using a parylene matrix chip, 25(OH)D 3 was modified through the nucleophilic addition of betaine aldehyde (BA) to form a hemiacetal salt. Such modified 25(OH)D 3 with a positive charge from BA could be effectively analyzed using MALDI-TOF mass spectrometry. Serum 25(OH)D 3 was extracted by liquid–liquid extraction (LLE) and quantified using MALDI-TOF mass spectrometry based on the parylene matrix chip. The intensity of the mass peak of 25(OH)D 3 was linearly correlated ( r 2 = 0.992) with the concentration of 25(OH)D 3 spiked in serum, and the LOD was 0.0056 pmol/μL. Energy drinks and vitamin D 3 tablets were also employed for the real sample analysis. Finally, the results of the chemiluminescence binding assay and MALDI-TOF mass spectrometry were statistically analyzed to determine the applicability of the method using the Bland–Altman test and Passing–Bablok regression.
更多
查看译文
关键词
MALDI-TOF mass spectrometry,Vitamin D,Parylene matrix chip,Serum analysis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要