Avoiding unconscious injection of vial-derived rubber particles during intra-articular drug administration

Osteoarthritis and Cartilage Open(2021)

引用 0|浏览2
暂无评分
摘要
This study shows the occurrence of rubber particles in 10-33% of the cases, when standard needles are used for the aspiration of drugs. We therefore recommend using industrially prefilled syringes, filter needles or removing the rubber stopper before withdrawing drugs from vials for intra-articular injections.
更多
查看译文
关键词
Vial coring,Intra-articular injections,Filter needle,Drug contamination,Foreign body reaction