A Dual-angle fiber dynamic light scattering system integrated with microfluidic chip for particle size measurement

Mei Chen, Yanghong Wang,Wei Li,Min Xia, Ziqiang Meng,Li Xia

Optics & Laser Technology(2022)

引用 8|浏览5
暂无评分
摘要
•A new fiber dynamic light scattering particle size measurement system with a microfluidic chip is proposed. The microfluidic analysis function and the optical function are integrated into an on-chip device, making the system miniaturized, highly integrated, and simple in optical alignment.•The design of ultra-short scattering light path and ultra-small sample volume effectively reduce the influence of multiple scattering on the measurement of high-concentration solutions, enabling the system to accurately measure polystyrene particle solution with a concentration of up to 14 mg/ml, with a relative measurement error of less than 6%.•The system realizes dual-angle dynamic light scattering measurement. The experimental results show that compared with single-angle measurement, dual-angle measurement has higher measurement stability (standard deviation less than 2.2% of the standard size).
更多
查看译文
关键词
Dynamic light scattering,Microfluidic chip,Particle size distribution,Multiple scattering