Very low contents (in the range of 10-9 g/g) of Ir in mantle-derived rock samples (komatiites) were nondestructively determined by INAA coupled with coincidence gamma-ray spectrometry using 16 Ge detectors. Aliquots of the same samples were analyzed by NiS fire-assay ICP-MS for Ir and other platinum group elements. Because the INAA procedure used in this study is non-destructive and is almost free from spectral interference in gamma-ray spectrometry, the INAA values of Ir contents obtained in this study can be highly reliable. Iridium values obtained by ICP-MS were consistent with the INAA values, implying that the ICP-MS values of Ir obtained in this study are equally reliable. Under the present experimental conditions, detection limits were estimated to be 1 pg/g, which corresponds to 0.1 pg for a sample mass of 0.1 g. These levels can be even lowered by an order of magnitude, if necessary, which cannot be achieved by ICP-MS carried out in this study. (c) 2021 Korean Nuclear Society, Published by Elsevier Korea LLC. This is an open access article under the CC BY-NC-ND license (http://creativecommons.org/licenses/by-nc-nd/4.0/).
更多
查看译文
关键词
Instrumental neutron activation analysis (INAA),Iridium,Ultra-basic rocks,Multiple Ge detector System,Coincidence counting,Inductively coupled plasma mass spectrometry (ICP-MS)