Mechanical Characteristics of the Flebogrif System-The New System of Mechano-Chemical Endovenous Ablation

MATERIALS(2022)

引用 0|浏览10
暂无评分
摘要
Non-thermal endovenous ablations, due to the lowest probability of complications, are the new method of treating chronic venous insufficiency-one of the most common diseases globally. The Flebogrif system (Balton Sp. z o.o., Warsaw, Poland) is a new mechano-chemical ablation system causing the mechanical damage of endothelium that allows for better sclerosant penetration into its wall. The purpose of the article is to provide mechanical characteristics in the form of force-displacement dependence for a single cutting element, and a bundle of cutting elements of Flebogrif as a whole for different levels of protrusion of the bundle of cutting elements. A TA.HD plus (Stable Micro Systems, Godalming, UK) analyzer equipped with special handles, was used for characteristics testing. The head movement speed used was 5 mm center dot s(-1). The Flebogrif system was tested for three cutting element protrusion levels: L = L-max, L = 0.9 center dot L-max, and L = 0.8 center dot L-max. Before testing, geometric measurement of the spacing of the cutting elements for three proposed protrusions was performed. It was established that decreasing the working length of the cutting elements will increase their rigidity, and, as a result, increase the force exerted on the internal surface of the vein wall. The obtained characteristics will allow for specifying contact force variability ranges and the corresponding diameter ranges of operated veins.
更多
查看译文
关键词
veins ablation, mechanical characteristics, non-thermal ablation, Flebogrif
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要