Inappropriate shock delivery as a result of electromagnetic interference originating from the faulty electrical installation

Milos D. Babic, Milosav Tomovic, Maja Milosevic, Branko Djurdjevic,Vasko Zugic,Aleksandra Nikolic

ANNALS OF NONINVASIVE ELECTROCARDIOLOGY(2022)

引用 2|浏览1
暂无评分
摘要
We present a case report of a 74-year-old male patient with an implantable cardioverter defibrillator who suffered an inappropriate defibrillation shock while bathing in the tub. Insight in the ICD stored electrogram episodes revealed electromagnetic interferences, with a typical 50 Hz electrical artifact mimicking fast ventricular tachycardia as a device misinterpreted. After this event, the maintenance workers investigated the electrical installation in the bathroom and revealed that there was voltage leaking between electrical installation and metal pipes. After the repair was completed without any additional programming, the patient has had no subsequent shocks.
更多
查看译文
关键词
electrical installation, electromagnetic interference, implantable cardioverter defibrillator, inappropriate shock
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要