Final report on supplementary comparison COOMET.PR-S7: Laser power responsivity (COOMET 599/RU/13)

Sergey A. Moskalyuk, Anatoly A. Liberman, Joshua Hadler,Paul Williams

METROLOGIA(2021)

引用 0|浏览1
暂无评分
摘要
Main text The National Institute of Standards and Technology (NIST), USA and the All-Russian Research Institute for Optical and Physical Measurements (VNIIOFI), Russia agreed in February 2013 to conduct a comparison on the laser power responsivity at wavelengths of 532 nm, 1.064 μm and 10.6 μm. The aim of this comparison is to assess the equivalence of the laser power responsivity between two laboratories. The comparison was conducted within the COOMET regional metrological organization (COOMET 599/RU/13) and was registered in the BIPM Key Comparison DataBase as a supplementary comparison with the identifier COOMET.PR-S7. The comparison was carried out using one detector head for measuring laser power. The detector head was supplied by VNIIOFI (OPHIR 10A). The results of this comparison essentially demonstrated agreement between the results obtained at the two participating laboratories. To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/ . The final report has been peer-reviewed and approved for publication by the CCPR, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).