Dosimetric study of the interplay effect using three-dimensional motion phantom in proton pencil beam scanning treatment of moving thoracic tumours

JOURNAL OF RADIOTHERAPY IN PRACTICE(2021)

引用 2|浏览1
暂无评分
摘要
Aim: The dosimetric and clinical advantages offered by implementation of pencil beam scanning (PBS) proton therapy for moving thoracic tumours is hindered by interplay effect. The purpose of this study is to evaluate the impact of large proton beam spot size along with adaptive aperture (AA) and various motion mitigation techniques on the interplay effect for a range of motion amplitudes in a three-dimensional (3D) respiratory motion phantom. Materials and Methods: Point doses using ionisation chamber (IC) and planner dose distributions with radiochromic film were compared against the corresponding treatment planning system (TPS) information. A 3D respiratory motion phantom was scanned either for static or 4D computed tomographic (CT) technique for 6-, 10- and 14-mm motion amplitudes in SI direction. For free breathing (FB) treatment, a tumour was contoured on maximum intensity projection scan and an average scan was used for treatment planning. Each FB treatment was delivered with one, three and five volumetric repaintings (VRs). Three phases (CT40-60%) were extracted from the 4D-CT scans of each motion amplitude for the respiratory-gated treatment and were used for the treatment planning and delivery. All treatment plans were made using AA and robustly optimised with 5-mm set-up and 3.5% density uncertainty. A total of 26 treatment plans were delivered to IC and film using static, dynamic and respiratory-gated treatments combinations. A percent dose difference between IC and TPS for the point dose and gamma indices for film-TPS planner dose comparison was used. Results: The dose profile of film and TPS for the static phantom matched well, and percent dose difference between IC and TPS was 0.4%. The percent dose difference for all the gated treatments were below 3.0% except 14-mm motion amplitude-gated treatment. The gamma passing rate was more than 95% for film-TPS comparison for all gated treatment for the investigated gamma acceptance criteria. For FB treatments, the percent dose difference for 6-, 10- and 14-mm motion amplitude was 1.4%, -2.7% and -4.1%, respectively. As the number of VR increased, the percent difference between measured and calculated values decreased. The gamma passing rate met the required tolerance for different acceptance criteria except for the 14-mm motion amplitude FB treatment. Conclusion: The PBS technique for the FB thoracic treatments up to 10-mm motion amplitude can be implemented with an acceptable accuracy using large proton beam spot size, AA and robust optimisation. The impact of the interplay effect can be reduced with VR and respiratory-gated treatment and extend the treatable tumour motion amplitude.
更多
查看译文
关键词
interplay effect, pencil beam scanning, phantom study, respiratory-gated treatment, volumetric repainting
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要