Calibration model and numerical simulation of a 7-hole probe based on pressure-velocity parameterization equation

MEASUREMENT SCIENCE AND TECHNOLOGY(2022)

引用 1|浏览3
暂无评分
摘要
As a sensor of a flow field detection system, a 7-hole probe can detect the flow field velocity and retrieve three-dimensional (3D) information of the flow field. Owing to its simple structure and strong environmental adaptability, it is particularly important to calibrate it when it is widely used in turbine machinery, aerospace, and other fields. To detect the 3D flow field in the middle atmosphere, a novel calibration method based on the potential flow theory is designed using a hemispherical 7-hole probe. The hemispherical 7-hole probe was numerically calibrated through numerical simulation, and the coefficients of the calibration equation are provided. In comparison with the traditional 7-hole probe calibration method, the calibration process is significantly shortened while maintaining good measurement accuracy. The velocity error was less than 5% and the angle error was approximately 0.5 degrees.
更多
查看译文
关键词
calibration, hemispherical 7-hole probe, numerical simulation, potential flow theory