B-PO03-001 IMPACT OF SENSING VECTOR ON INSERTABLE CARDIAC MONITOR DETECTION OF P WAVE AMPLITUDE: IMPLICATION FOR IMPROVED ATRIAL ARRHYTHMIA DIAGNOSIS

Heart Rhythm(2021)

引用 1|浏览6
暂无评分
摘要
P-wave visibility is important for diagnosis of atrial arrhythmias via insertable cardiac monitors (ICM). To compare the sensing capability of ICM with different vector length: Medtronic REVEAL LINQ (45mm) vs. Biotronik Biomonitor BMIII (75mm). We accessed de-identified remote monitoring transmissions from n=40 patients with BMIII from Biotronik for comparison with n=80 gender- and body mass index (BMI)-matched patients with REVEAL LINQ. Digital measurement of calibrated ICM electrograms was performed using DigitizeIt© (V2.3.3, Germany) of 5 consecutive P-QRS complexes by 3 investigators independently. P-wave amplitude was measured from baseline to peak and R-wave amplitude was measured from peak-to-peak. Further, we evaluated the impact of BMI on P-wave visibility. Patients in both groups are well matched for gender and BMI (53% male, mean BMI 26.7 kg/m2; both p=NS). P-wave amplitude was 89% larger in the BMIII vs. LINQ (0.072±0.039 vs. 0.038±0.022mV, p=0.0001) and R-wave amplitude was 20% larger although this did not reach statistical significance (0.83±0.42 vs. 0.69±0.45mV, p=0.095). Multivariate regression analysis showed that increasing BMI (figure), female gender and LINQ were independently associated with lower P-wave amplitude (all P<0.001). Inter-operator reproducibility of P-wave measures was very good (ICC 0.91, P<0.001). The longer ICM sensing vector of BMIII yielded considerable larger P-wave amplitude that was not impacted by increasing BMI. Better P-wave visibility may improve clinical decision-making after device detected arrhythmias.
更多
查看译文
关键词
Arrhythmia Detection,Physiological Monitoring
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要