Paradoxical pacing spikes during a wide complex tachycardia in biventricular defibrillator. What is the mechanism? Is it undersensing due to troubleshooting?

JOURNAL OF ARRHYTHMIA(2022)

引用 0|浏览0
暂无评分
摘要
We present the case of a 68-year-old man with a cardiac resynchronization therapy-defibrillator Medtronic device implanted 3 years ago, admitted to our hospital due to a wide complex tachycardia with unexpected pacing spikes inside QRS complexes. Although that could be easily attributed to ventricular undersensing, more complicated mechanisms are involved, including the origin of the tachycardia, the parameters of the device, and the position of the device and the leads.
更多
查看译文
关键词
defibrillator function, paradoxical pacing, resynchronization therapy, wide complex tachycardia