AGILE Instrument: Advanced Energetic Ion Electron Telescope

IEEE Transactions on Nuclear Science(2022)

引用 0|浏览10
暂无评分
摘要
Identifying charged particles using crucial features of the reconstructed waveforms is a widely exploited strategy for signal analysis in the class of pulse shape discrimination (PSD) techniques. This document details the design, development, and characterization of the hardware of a compact and low-power consumption telescope for data collection in space. The advanced energetic ion electron telescope (AGILE) experimental apparatus includes a stack of three 20-mm-diameter circular Si detector layers read-out by custom-made double-gain electronics that is optimized for the wide spectrum of expected energy depositions (1–100 MeV/nucleon). At the end of its front-end chain, the instrument uses a fast digitizing and sampler device based on commercial components, integrated in a control board developed by the collaboration. In this report, we motivate AGILE’s hardware design and report the results obtained during the instrument’s prototype laboratory test characterization.
更多
查看译文
关键词
Digital signal processing,particle measurements,silicon radiation detectors,space applications