Design of a DT neutron source based PGNAA facility for element determination in aqueous solution

Applied Radiation and Isotopes(2022)

引用 0|浏览3
暂无评分
摘要
A DT neutron source-based prompt gamma ray neutron activation (PGNAA) facility for bulk sample analysis was designed and developed in this research. The aqueous samples containing chlorine and boron with known concentration were determined for the calibration curve by the facility. The neutron self-shielding effect was corrected by internal standard method. The minimum detectable concentration of boron and chlorine are 1.37 ± 0.42 and 12.51 ± 3.80 mg/L, respectively. Finally, five mixture aqueous solution samples were measured for the facility performance test. The maximum relative deviation of boron and chlorine are 4.49% and 5.32% respectively.
更多
查看译文
关键词
PGNAA,MCNP,Neutron self-shielding,DT neutron Source,Boron and chlorine,MDC