谷歌浏览器插件
订阅小程序
在清言上使用

Isotope Ratio Outlier Analysis (IROA) for HPLC-TOFMS-Based Metabolomics of Human Urine.

Metabolites(2022)

引用 0|浏览16
暂无评分
摘要
Metabolic fingerprinting by mass spectrometry aims at the comprehensive, semiquantitative analysis of metabolites. Isotope dilution, if successfully implemented, may provide a more reliable, relative quantification. Therefore, the 13C labeled yeast extract of the IROA TruQuant kit was added as an internal standard (IS) to human urine samples measured in full-scan mode on a high-performance liquid chromatography-time-of-flight mass spectrometer (HPLC-TOFMS) system. The isotope ratio approach enabled the analysis of 112 metabolites. The correlation with reference data did not improve significantly using 12C/13C ratios compared to absolute 12C peak areas. Moreover, using an intricate 13C-labeled standard increased the complexity of the mass spectra, which made correct signal annotation more challenging. On the positive side, the ratio approach helps to reduce batch effects, but it does not perform better than computational methods such as the "removebatcheffect" function in the R package Limma.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要