Impact of Chocolate percutaneous transluminal angioplasty balloon on vessel preparation in drug-coated balloon angioplasty for femoropopliteal lesion

CVIR Endovascular(2022)

Cited 1|Views12
No score
Abstract
Purpose To compare the impact of Chocolate and conventional balloons on vessel preparation in percutaneous transluminal angioplasty. Materials and methods This single-center retrospective study included 111 endovascular therapy consecutive cases of femoropopliteal lesions using drug-coated balloon strategy with a 1:1 pre-dilation balloon diameter between February 2020 and August 2021, divided into the Chocolate percutaneous transluminal angioplasty ( n = 48) and conventional ( n = 63) groups. Before the availability of Chocolate balloons in Japan (December 2020), a standard semi-compliant or non-compliant balloon was used for vessel preparation. The primary endpoint was rate of severe dissection after pre-dilatation. Secondary endpoints were angiographic percent diameter stenosis, bailout stent rate, primary patency rate, and freedom from target-lesion-revascularization rate at six months. Results There was no significant difference in patient and lesion characteristics. The procedural characteristics comprised balloon length 90 ± 37 and 149 ± 95 mm ( P = 0.004) and inflation pressure 11 ± 3 and 16 ± 7 atm ( P < 0.001) in the Chocolate and conventional groups, respectively. Regarding primary endpoint, rates of severe dissection were 4.2% and 25% ( P = 0.003); regarding secondary endpoints, percent diameter stenosis was 18 ± 15% and 20 ± 17% ( P = 0.409), and the rate of bailout stenting was 2.1% and 15.9% ( P = 0.016) in the Chocolate and conventional groups, respectively. The primary patency rates at six months were 89.1% and 85.2% ( P = 0.670), and freedom from target-lesion-revascularization rate at six months was 100% and 92.8% ( P = 0.691) in the Chocolate and conventional groups, respectively. Conclusion Chocolate percutaneous transluminal angioplasty balloons reduce the rate of severe dissection while maintaining a sufficient dilatation effect during drug-coated balloon vessel preparation.
More
Translated text
Key words
Peripheral artery disease, Endovascular procedures, Femoral artery, Popliteal artery, Drug-coated balloon
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined