Closed-Loop Control of Arterial CO2 in Mechanical Ventilation of Neonates.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)(2022)

引用 0|浏览4
暂无评分
摘要
During mechanical ventilation of the neonate the main goal is to stabilize respiratory function of the often premature lungs. Ventilating the patient without inflicting harm is then the subordinated next goal. Ideally the arterial partial pressure of CO lays within a normocapnic range and fluctuations are kept minimal. By closely monitoring CO and controlling ventilation parameters accordingly, CO levels in the blood can be managed. We present an approach consisting of a cascaded controller for arterial CO by approximating arterial partial pressure PaCO from end-tidal PetCO. As a proof of concept, feasibility of the controller was first evaluated on a mathematical patient model and subsequently in-vivo in lamb experiments. The controller is able to regulate CO into a normocapnic range in both setups with satisfactory stationarity within the target range. Estimation of the arterial partial pressure of CO remains a critical aspect that needs to be further investigated. Clinical relevance-Closed-loop control of CO2 in mechanical ventilation aims to avoid PaC O2 extremes and to reduce fluctuations. Both are a relevant risk factors especially for neurological complications among preterm newborns.
更多
查看译文
关键词
arterial control,mechanical ventilation,neonates,closed-loop