Evaluating the Empatica E4 Derived Heart Rate and Heart Rate Variability Measures in Older Men and Women.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC)(2022)

引用 1|浏览9
暂无评分
摘要
Wearable heart rate monitors offer a cost-effective way of non-invasive, long-term monitoring of cardiac health. Validation of wearable technologies in an older populations is essential for evaluating their effectiveness during deployment in healthcare settings. To this end, we evaluated the validity of heart rate measures from a wearable device, Empatica E4, and compared them to the electrocardiography (ECG). We collected E4 data simultaneously with ECG in thirty-five older men and women during an overnight sleep recording in the laboratory. We propose a robust approach to resolve the missing inter-beat interval (IBI) data and improve the quality of E4 derived measures. We also evaluated the concordance of heart rate (HR) and heart rate variability (HRV) measures with ECG. The results demonstrate that the automatic E4 heart rate measures capture long-term HRV whilst the short-term metrics are affected by missing IBIs. Our approach provides an effective way to resolve the missing IBI issue of E4 and extracts reliable heart rate measures that are concordant with ECG. Clinical Relevance- This work discusses data quality challenges in heart rate data acquired by wearables and provides an efficient and reliable approach for extracting heart rate measures from the E4 wearable device and validates the metrics in older adults.