Treating posteriorly eroded glenoids with augmented baseplate or bony increased-offset reverse shoulder arthroplasty: a finite element comparison

Asees Kaur,Carson Studders, Delaney Haugan, David Saliken,Joshua W. Giles

Seminars in Arthroplasty: JSES(2023)

引用 0|浏览0
暂无评分
摘要
Augmented reverse shoulder arthroplasty (RSA) implants restore glenohumeral joint alignment in cases of asymmetric glenoid wear. However, no consensus has been reached on whether the use of metallic augmented RSA baseplates and bone graft reconstruction are equivalent in terms of implant fixation and risk of implant loosening. Therefore, the purpose of this study was to compare two augmented RSA designs by assessing the amount of interfacial micromotion generated under realistic physiological loading. Finite element analysis (FEA) models of 9 scapulae with Walch-type B2 or B3 glenoid morphology were virtually implanted with both a metallic augmented baseplate (AUG-RSA) and using the angled bony increased offset RSA procedure (BIO-RSA). Simulation of physiological loading was performed on each of the 18 FEA models. The relative tangential and normal micromotion at the implant-to-glenoid interface was compared in each anatomical quadrant. The AUG-RSA and angled BIO-RSA showed similar magnitudes of micromotion in most anatomical quadrants of the glenoid. Within the superior quadrant, AUG-RSA displayed a higher magnitude of mean and maximum tangential micromotion (mean: 16.6 ± 2.4 μm, p<0.000; max: 35.1 ± 5.3 μm, p<0.000). The proportion of the posterior quadrant experiencing greater than 50 microns of micromotion was also statistically greater with AUG-RSA (5.8 ± 2.5 %, p=0.047). Due to its statistically greater micromotion and portions of contact exceeding the accepted 50 micron threshold, the AUG-RSA may be more likely to have inhibited bone on-growth. However, the clinical importance of these differences remains unclear.
更多
查看译文
关键词
Basic Science Study, Finite Element Analysis
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要