Capacitive coupling leading to electrical skin burn injury during laparoscopic surgery.

Woo Jun Kim,Gyung Mo Son,In Young Lee, Sung Uk Yun, Gye Rok Jeon,Dong-Hoon Shin, Myung Sook Kwon, Jae Yeong Kwak,Kwang-Ryul Baek

Journal of minimally invasive surgery(2022)

引用 0|浏览2
暂无评分
摘要
This study demonstrated that even with a plastic trocar and electrosurgical instruments that have intact insulation, if an open circuit is formed, capacitive coupling increases, and trocar-site burn can occur. When using electrocautery, careful manipulation must be taken to avoid creating an open circuit to prevent capacitive coupling related to electrical skin burn.