Noise Performance of the Ganymed Laser Altimeter Receiver

T. Behnke, R. Kallenbach, M. Fuji,M. Kobayashi, Y. Sato, K. Enya,C. Huttig,A. Stark, S. DelTogno, C. Althaus

2022 ESA Workshop on Aerospace EMC (Aerospace EMC)(2022)

引用 0|浏览3
暂无评分
摘要
We report here on qualification and calibration test results demonstrating the science performance of the Avalanche Photodiode (APD) detector of the Ganymede Laser Altimeter (GALA, Fig. 1). Most importantly, the measures for noise reduction of the analogue receiver electronics are described. This includes proper grounding of the APD module (EXCELITAS C30654) and filtering in its vicinity and the additional measures to isolate switching noise of the analogue-to-digital convertors (ADCs) from the sensitive single-ended analogue sensor pre-amplifier electronics of the APD. Tests with realistic electron irradiation doses being representative for the environment of the Jovian magnetosphere and additional shielding measures have been performed in order to ensure that the dark-current noise level of the receiver does not increase significantly over the mission time.
更多
查看译文
关键词
noise performance,ganymed Laser Altimeter receiver,calibration test results,science performance,Avalanche Photodiode detector,Ganymede Laser Altimeter,GALA,Fig,noise reduction,analogue receiver electronics,proper grounding,APD module,EXCELITAS C30654,additional measures,switching noise,analogue-to-digital convertors,sensitive single-ended analogue sensor pre-amplifier electronics,realistic electron irradiation,additional shielding measures,dark-current noise level