Total Body Irradiation for Standard Treatment Rooms: a Robust Sweeping Beam Technique with Respect to the Body Shape.

Reports of Practical Oncology and Radiotherapy(2022)

引用 27|浏览9
暂无评分
摘要
Background:The purpose of this work is to improve a sweeping beam technique for total body irradiation (TBI) on a low flat couch using a varying patient thickness model. We designed a flat couch for total body irradiation in supine and prone position. Three generic arcs with rectangular segments for a patient torso thickness of 16, 22 and 28 cm were generated with respect to varying patient thickness of four particular parts of the body: head, torso, thighs and calves.Materials and methods:Longitudinal and transversal dose profiles were measured using an ionization chamber and the EBT3 gafchromic film in a solid water slab phantom. The robustness of the method was examined in phantoms of different thicknesses.Results:Measured dose homogeneity stays within ±10% of prescribed dose for all of the three patient thickness models. The robustness of the method was evaluated as the increase in dose in the phantom center of 0.7% per 1 cm reduction in phantom thickness.Conclusion:The method is applicable for the broad range of patient sizes, comfortable for patients, robust and suitable for standard treatment rooms with a standard linear accelerator. It requires minimal investments into equipment.
更多
查看译文
关键词
total body irradiation,TBI,arc therapy,sweeping beam technique,patient comfort