Introduction Worldwide more and more primary knee replacements are being performed. Kinematic alignment (KA) as one of many methods of surgical alignment has been shown to have a significant impact on kinematics and function. The aim of the present study was to compare KA and mechanical alignment (MA) with regard to femorotibial kinematics. Materials and methods Eight fresh frozen human specimens were tested on a knee rig during active knee flexion from 30 to 130°. Within the same specimen a medial stabilized (MS) implant design was used first with KA and then with MA. Results The femorotibial kinematics showed more internal rotation of the tibia in KA compared to MA. At the same time, there was a larger medial rotation point in KA. Both alignment methods showed femoral rollback over the knee bend. Conclusion Relating to an increased internal rotation and a more precise medial pivot point, it can be concluded that KA combined with a MS implant design may partially support the reproduction of physiological knee joint mechanics.