Real-Time Temperature Rise Estimation during Irreversible Electroporation Treatment through State-Space Modeling

BIOENGINEERING-BASEL(2022)

引用 4|浏览9
暂无评分
摘要
To evaluate the feasibility of real-time temperature monitoring during an electroporation-based therapy procedure, a data-driven state-space model was developed. Agar phantoms mimicking low conductivity (LC) and high conductivity (HC) tissues were tested under the influences of high (HV) and low (LV) applied voltages. Real-time changes in impedance, measured by Fourier Analysis SpecTroscopy (FAST) along with the known tissue conductivity and applied voltages, were used to train the model. A theoretical finite element model was used for external validation of the model, producing model fits of 95.8, 88.4, 90.7, and 93.7% at 4 mm and 93.2, 58.9, 90.0, and 90.1% at 10 mm for the HV-HC, LV-LC, HV-LC, and LV-HC groups, respectively. The proposed model suggests that real-time temperature monitoring may be achieved with good accuracy through the use of real-time impedance monitoring.
更多
查看译文
关键词
pulsed field ablation, PFA, electroporation, H-FIRE, thermal mitigation, temperature prediction, black-box modeling, agar phantom