The impact of silver on uranium quantification by kinetic phosphorescence analysis

Ean Arnold, Nic Uhnak, Jenna Schroeder,Lori Metz,Chelsie Beck

JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY(2022)

引用 0|浏览4
暂无评分
摘要
The quantification of uranium (U) by Kinetic Phosphorescence Analysis (KPA) is negatively impacted by the presence of silver (Ag). KPA can be combined with gamma spectroscopy to determine recovery of short-lived uranium isotopes after radiochemical purification in nuclear forensics relevant samples. Fission products can impact accurate gamma spectroscopy analysis. Removal of the 132 Te fission product can be accomplished via deposition on a Ag disk. The purpose of this work is to determine the concentration of Ag that begins to quench the phosphorescent signal as part of sample purification scheme for U and the necessary steps for successful sample analysis via gamma spectroscopy and KPA.
更多
查看译文
关键词
Kinetic phosphorescence analysis,Uranium analysis,Nuclear forensics,Tellurium,Silver