Effect of Calibration for Tissue Differentiation Between Healthy and Neoplasm Lung Using Minimally Invasive Electrical Impedance Spectroscopy

IEEE ACCESS(2022)

引用 2|浏览4
暂无评分
摘要
This study proposes a calibration method and analyses the effect of this calibration in lung measures, using minimally invasive electrical impedance spectroscopy with the 3-electrode method, for tissue differentiation between healthy and neoplasm lung tissue. Tissue measurements were performed in 99 patients [54 healthy tissue and 15 neoplastic tissue samples obtained] with an indicated bronchoscopy. Statistically significant difference (P < 0.001) were found between healthy lung tissue and neoplasm lung tissue in bioimpedance parameters. The calibration of the bioimpedance measures with respect to a measure performed in bronchi reduces the inter-patient dispersion, increasing the sensitivity, decreasing the specificity and increasing the area below the ROC curve for three out of four impedance-derived estimators. Results also show that there are no significant differences between healthy lung tissue among smoker, non-smoker and ex-smoker samples, which was initially stated as a possible cause of EIS measurement dispersion in lungs.
更多
查看译文
关键词
Lungs, Impedance, Bioimpedance, Impedance measurement, Current measurement, Frequency measurement, Electrodes, Spectroscopy, Calibration, Respiratory system, Bronchi, bronchoscopy, calibration, electrical impedance spectroscopy (EIS), lung