Novel findings of an old algorithm: PVC response "Atrial Pace'' initiates atrial arrhythmias in Abbott ICD and CRT-D devices.
Pacing and clinical electrophysiology : PACE(2023)
摘要
The A-Pace on PVC setting was frequently used in our patients. Our study shows that a considerable number of patients had at least one AHRE triggered by the algorithm. The use of this algorithm should be carefully reconsidered.
更多查看译文
关键词
algorithm,atrial tachycardia,premature ventricular contractions,trigger