The analysis of the National Ignition Facility (NIF) neutron time-of-flight (nToF) detectors uses a forward-fit routine that depends critically on the instrument response functions (IRFs) of the diagnostics. The details of the IRFs used can have large impacts on measurements such as ion temperature and down-scattered ratio (DSR). Here, we report on the recent steps taken to construct and validate nToF IRFs at the NIF to an increased degree of accuracy, as well as remove the need for fixed DSR baseline offsets. The IRF is treated in two parts: a “core,” measured experimentally with an x-ray impulse source, and a “tail” that occurs later in time and has limited experimental data. The tail region is calibrated with the data from indirect drive exploding pusher shots, which have little neutron scattering and are traditionally assumed to have zero DSR. Using analytic modeling estimates, the non-zero DSR for these shots is estimated. The impact of varying IRF tail components on DSR is investigated with a systematic parameter study, and good agreement is found with the non-zero DSR estimates. These approaches will be used to improve the precision and uncertainty of NIF nToF DSR measurements.