Technical Note: Water-equivalent Thickness of Superflab Bolus Material at Diagnostic X-Ray Energies.

Medical physics on CD-ROM/Medical physics(2023)

Cited 0|Views7
No score
Abstract
PurposeThe purpose of this work was to determine the water-equivalent thickness of Superflab bolus material for narrow and broad field-of-view (FOV) x-ray geometries at diagnostic x-ray energies. MethodsTransmission measurements were performed for incremental thicknesses of Superflab bolus material and water in narrow and broad FOV x-ray geometries. The transmission data was fit to a non-linear model for x-ray transmission - the Archer model. Water-equivalent thickness of Superflab was calculated based upon fitting parameters to transmission curves for 75, 95, and 115 kV x-ray tube voltages. Measured x-ray transmission factors for water and Superflab were used to determine the water equivalence of Superflab. ResultsFor all x-ray tube voltages and geometries, the water equivalence of Superflab was greater than one, indicating that Superflab is more attenuating than water. This effect was stronger for broad FOV geometries. At 95 kV, 30 cm of Superflab corresponded to 32.0 cm of water in the narrow FOV geometry, and 34.3 cm of water in the broad FOV geometry. The Archer model fitting parameters and Superflab water equivalence are reported for all x-ray beam conditions explored in this work. ConclusionsSuperflab bolus material is more attenuating than water at diagnostic x-ray energies. The Archer model and its respective fitting parameters reported in this work may be used to estimate the water-equivalent thickness of Superflab for diagnostic x-ray spectra.
More
Translated text
Key words
broad-beam attenuation,diagnostic x-ray imaging,narrow-beam attenuation,Superflab,water-equivalent thickness
AI Read Science
Must-Reading Tree
Example
Generate MRT to find the research sequence of this paper
Chat Paper
Summary is being generated by the instructions you defined