Analytical method for calculating internal stray radiation from an IR spectrometer based on the view factor

Applied optics(2023)

引用 1|浏览10
暂无评分
摘要
Optomechanical components such as the lens barrels and frames of IR spectrometers produce strong internal stray radiation, which reduces the instrument's SNR and dynamic range. An IR internal stray radiation calculation method based on an analytical model of the view factor is proposed. The mathematical model of the view factor calculation method of typical optomechanical components is established. For any IR optical systems, the internal stray radiation can be quickly and accurately calculated by adjusting the coordinate systems in the calculation method. Based on the proposed method, the internal stray radiation of a double-pass long-wave IR spectrometer was calculated. The calculation results are consistent with the simulation results. The RMS value of the relative error between the calculated value and the simulated value is around 11%. To verify the proposed method, an experiment was conducted to test the internal stray radiation of the long-wave IR spectrometer. The internal stray radiation test results agree with the calculated and simulated results, and the relative error between the test results and the calculation results is within 9%. (c) 2023 Optica Publishing Group
更多
查看译文
关键词
ir spectrometer,internal stray radiation,analytical method
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper