Effect of surgical microscope and illuminated chopper on anterior chamber temperature

BMC Ophthalmology(2023)

引用 2|浏览8
暂无评分
摘要
Background To evaluate the effect of the light intensity of the surgical microscope and illuminated chopper on the anterior chamber temperature. Study design Experimental study. Methods A model eye (Kitaro WetLab System; FCI Ophthalmics, Pembroke, MA, USA) was used in this experimental study. The illuminance of a surgical microscope (Leica M300; Leica Microsystems, Wetzlar, Germany) and illuminated chopper (iChopper NAM-25 GB; Oculight, Korea) with a light source (iVision; Oculight) was measured using an illuminometer. In addition, the temperature in the anterior chamber of the model eye filled with balanced salt solution when using the surgical microscope with a light intensity from level 1 to level 6 and the illuminated chopper at 99% light intensity was measured for 10 min. Results The anterior chamber temperature was increased by 0.2, 0.5, 1.0, and 1.4 ℃ when using the surgical microscope at level 3 (10050 lux), 4 (16490 lux), 5 (24900 lux), and 6 (32500 lux), respectively, for 10 min. The illuminated chopper at 99% light intensity (14893 lux) positioned in the anterior chamber increased the anterior chamber temperature by 0.2° C after 10 min, which was equal to the increase in the temperature caused by the surgical microscope at level 3. Conclusion The photothermal effect of the illuminated chopper directly positioned in the anterior chamber appeared to be similar to that of a microscope with similar illuminance. Therefore, the illuminated chopper is safe in terms of anterior chamber temperature changes in cataract surgery.
更多
查看译文
关键词
Microscope,Illuminated chopper,Temperature
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要