Identification of a frit-related sample carryover in newborn screening by tandem mass spectrometry

Kyunghun Kim,Howon Lee,Jeong Joong Lee, Kyoungho Cha, Nam Hyun Park, Young Keun Shin,Hyojin Chae,Eun-Jee Oh

Journal of Mass Spectrometry and Advances in the Clinical Lab(2023)

引用 0|浏览10
暂无评分
摘要
•Sample carryover in newborn screening test can increase false positive results.•The risk of carryover is increased with newer generations of mass spectrometers.•Column-related carryover due to a stainless-steel frit inline filter was detected.•Compound-dependent factor of stainless-steel frit-related carryover is suggested.•Routine backpressure monitoring and timely replacement of frit rings are mandatory.
更多
查看译文
关键词
DBS,ESI,FIA,FTN,H,HPLC,IEM,IPA,L,LC,LLOQ,MS,MS/MS,NBS,PEEK,pI,QC,SM,TIC,UPLC
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要