Analysis of Various Modalities for Intraocular Pressure Measurement in Relation to Keratoconus Severity in 246 Eyes of the Homburg Keratoconus Center.

Cornea(2023)

引用 0|浏览0
暂无评分
摘要
CST-based bIOP seems to be best suited for IOP measurement in KC because it did not correlate with TCT in contrast to IOP measurements by ORA or GAT. The measurement results from GAT in patients with KC should be interpreted with care and always in view of corneal thickness. As a thumb rule, we suggest to add at least 2 mm Hg to the measured GAT value.
更多
查看译文
关键词
intraocular pressure measurement,keratoconus severity,homburg keratoconus center,eyes