A small-sized fluorescence detector (referred to as a pipette tip [PT]-reader) was developed for a pipette tip-based biosensor. The PT-reader allows us to measure the fluorescence intensity of a solution in a truncated cone-shaped pipette tip with only the tip inserted into the PT-reader. A pipette holder made from a mixture of polydimethylsiloxane (PDMS) and carbon black was capable of the rigorous position arrangement of a truncated cone shaped-pipette tip and the prevention of stray light. The detection performance of the PT-reader was evaluated by measurement of resorufin. The limit of detection (LOD; 3σ) and the relative standard deviation (RSD, n = 4) were estimated to be 0.46 μM and 0.47–4.1%, respectively. This performance was comparable to that of a desktop-type fluorescence microplate reader. In addition, the PT-reader was applied to the quantification of immunoglobulin A (IgA), and the LOD (3σ) of IgA was estimated to be 1.0 ng/mL. The quantitation values of IgA in human saliva obtained by the PT-based enzyme-linked immunosorbent assay (PT-ELISA) were in agreement with those obtained by conventional ELISA. The PT-reader is expected to be useful for low-cost and user-friendly measurements, and the technique of device development proposed in this study will contribute to the progress of on-site medical diagnosis.
更多
查看译文
关键词
Pipette tip-based biosensor,Fluorescence detector,ELISA,IgA,On-site diagnosis