Neurospin is a neuroscience research center located in France at the Atomic Energy Commission (CEA Saclay). The facility is hosting an innovative whole-body 11.7 T MRI system that has delivered its first images in October 2021. The core part of the Iseult MRI is an actively shielded NbTi magnet cooled with a pressurized superfluid helium bath at 0.125 MPa and 1.8 K, providing a homogeneous magnetic field of 11.7 T within a 90 cm warm bore. After nearly twenty years of work and efforts, the magnet successfully reached its nominal field for the first time in July 2019. The field homogeneity has been adjusted and the control system tested against internal and external faults that could affect the future MRI operation. MRI peripheral equipment has been integrated and interactions between the gradient coils and the magnet and their impact on cryogenics and on the magnet safety system have also been studied. The MR scanner is now kept permanently at nominal field and the final calibration is on-going to prepare the first acquisition on a human volunteer. The paper will present the Iseult MRI commissioning status and the first images obtained, as well as a first feedback on the cryogenic plant operation after three years and a half at 1.8 K.