Micro-Computed Tomography Beamline of the Australian Synchrotron: Micron-Size Spatial Resolution X-ray Imaging

Applied sciences(2023)

引用 2|浏览23
暂无评分
摘要
The first new beamline of the BRIGHT project—involving the construction of eight new beamlines at the Australian Synchrotron—is the Micro-Computed Tomography (MCT) beamline. MCT will extend the facility’s capability for higher spatial resolution X-ray-computed tomographic imaging allowing for commensurately smaller samples in comparison with the existing Imaging and Medical Beamline (IMBL). The source is a bending-magnet and it is operating in the X-ray energy range from 8 to 40 keV. The beamline provides important new capability for a range of biological and material-science applications. Several imaging modes will be offered such as various X-ray phase-contrast modalities (propagation-based, grating-based, and speckle-based), in addition to conventional absorption contrast. The unique properties of synchrotron radiation sources (high coherence, energy tunability, and high brightness) are predominantly well-suited for producing phase contrast data. An update on the progress of the MCT project in delivering high-spatial-resolution imaging (in the order of micron size) of mm-scale objects will be presented in detail with some imaging results from the hot-commissioning stage.
更多
查看译文
关键词
X-ray micro-CT,absorption contrast,phase-contrast,computed tomography,synchrotron