Accuracy of pilot balloon palpation for cuff pressure assessment in small versus large sized tubes: a prospective non-randomized observational study

Scientific Reports(2023)

引用 0|浏览2
暂无评分
摘要
Pilot balloon palpation is still a commonly used method to evaluate cuff pressure of the endotracheal tube after intubation. This study determined whether the size of the tracheal tube influenced the accuracy of pilot balloon palpation. A prospective observational analysis of 208 patients intubated with an endotracheal tube of internal diameter (ID) 6.0 or 8.0 was conducted. An anesthesiologist judged the cuff pressure by manual pilot balloon palpation, and then measured the cuff pressure with a pressure gauge. Cuff pressure exceeding 20–30 cmH 2 O was defined as false recognition. The intracuff pressure was significantly higher in ID 6.0 tube than in the ID 8.0 tube (41.9 ± 18.8 cmH 2 O vs. 30.3 ± 11.9 cmH 2 O, p < 0.001). The number of patients that were mistakenly perceived to have appropriate cuff pressure by pilot balloon palpation was significantly higher in the ID 6.0 group compared to the ID 8.0 group (85 (81.7%) vs. 64 (61.5%), p = 0.001). Therefore, a smaller tube size may further increase risk of inaccurate measurement by pilot balloon palpation and although pressure gauge is recommended for all sizes to maximize accuracy, groups with increased risk factors should be targeted for standardized use of the pressure gauge.
更多
查看译文
关键词
Outcomes research,Respiratory signs and symptoms,Risk factors,Science,Humanities and Social Sciences,multidisciplinary
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要