Validation of a Carbon Monoxide Sampling Method Using Polypropylene Syringes: a Reliable Alternative Method, Easy to Implement in the Field or Emergency Situations

Béatrice Le Roy, Lucien Valente, Gérald Provini,Jean-Ulrich Mullot

Environmental monitoring and assessment(2023)

引用 0|浏览1
暂无评分
摘要
Carbon monoxide (CO) is a lethal gas, present during incomplete combustion of carbonaceous materials. CO may be present in certain occupational atmospheres or during accidental events such as fires. Colorless and odorless, its presence can only be detected analytically. Nevertheless, the measurement methods available today may either be lacking (not available, for example, for military people deployed in the field) or not completely adapted (interference for electrochemical detectors, cost for infrared detectors). Another solution is to take samples on the field and then analyze them in a laboratory. Tedlar® bags or canisters can be used for this purpose but are relatively cumbersome. An alternative, not well described in the literature and not metrologically validated, consists in using plastic syringes. In order to generalize the use of this system and to characterize it in terms of performance and stability over time, we conducted a validation study. This method was validated using a 50-cc polypropylene syringe, over a concentration range of 2 to 40 ppm. The sampling system is efficient (sample yields between 101 and 102
更多
查看译文
关键词
Carbon monoxide,Sampling,Polypropylene syringe,Validation