Validity and reliability of integrated pressure level real-ear-to-coupler difference measurements

INTERNATIONAL JOURNAL OF AUDIOLOGY(2023)

引用 1|浏览5
暂无评分
摘要
Objectives(1) To validate the measurement of foam-tip real-ear-to-coupler differences (wRECD) using an integrated pressure level (IPL) method and (2) to compare the reliability of this method to SPL-based measurement of the wRECD.DesignSPL-based wRECD and the proposed IPL wRECD measurement were completed bilaterally. Test-retest reliability of IPL wRECD was determined with full re-insertion into the ear canal and compared to published SPL wRECD test-retest data.Study sample22 adults with normal hearing and middle ear status were recruited.ResultsDifferences between SPL-based wRECD and IPL wRECD measurements were within 1.51 dB on average below 5000 Hz. At and above 5000 Hz, IPL wRECD exceeded SPL wRECDs by 6.11 dB on average. The average test-retest difference for IPL wRECD across all assessed frequencies was 0.75 dB with the greatest improvements in reliability found below 750 Hz and above 3000 Hz.ConclusionsIPL wRECD yielded improved estimates compared to SPL wRECD in high frequencies, where standing-wave interference is present. Independence from standing wave interference resulted in increased wRECD values above 4000 Hz using the IPL measurement paradigm. IPL wRECD is more reliable than SPL wRECD, does not require precise probe-microphone placement, and provides a wider valid wRECD bandwidth than SPL-based measurement.
更多
查看译文
关键词
pressure level,reliability,real-ear-to-coupler