Resolve is a payload hosting an x-ray microcalorimeter detector operated at 50 mK in the x-ray imaging and spectroscopy mission. It is currently under development as part of an international collaboration and is planned to be launched in 2023. A primary technical concern is the microvibration interference in the sensitive microcalorimeter detector caused by the spacecraft bus components. We conducted a series of verification tests in 2021 to 2022 on the ground, the results of which are reported here. We defined the microvibration interface between the spacecraft and the Resolve instrument. In the instrument-level test, the flight-model hardware was tested against the interface level by injecting it with microvibrations and evaluating the instrument response using the 50 mK stage temperature stability, adiabatic demagnetization refrigerator magnet current consumption rate, and detector noise spectra. We found strong responses when injecting microvibration at similar to 200, 380, and 610 Hz. In the former two cases, the beat between the injected frequency and cryocooler frequency harmonics were observed in the detector noise spectra. In the spacecraft-level test, the acceleration and instrument responses were measured with and without suspension of the entire spacecraft. The reaction wheels (RWs) and inertial reference units (IRUs), two major sources of microvibration among the bus components, were operated. In conclusion, the observed responses of Resolve are within the acceptable levels in the nominal operational range of the RWs and IRUs. There is no evidence that the resultant energy resolution degradation is beyond the current allocation of noise budget.
更多
查看译文
关键词
microvibration interference,spectroscopy mission,x-ray
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要