Range margin reduction in carbon ion therapy: potential benefits of using radioactive ion beams

Research Square (Research Square)(2022)

引用 0|浏览0
暂无评分
摘要
Abstract Sharp dose gradients and high biological effectiveness make heavy ions such as 12C an ideal tool to treat deep-seated tumors, however, at the same time, sensitive to errors in the range prediction. Tumor safety margins mitigate these uncertainties, but during the irradiation they lead to unavoidable damage to the surrounding healthy tissue. To fully exploit the Bragg peak benefits, a large effort is put into establishing precise range verification methods. Despite positron emission tomography being widely in use for this purpose in 12C therapy, the low count rates, biological washout, and broad activity distribution still limit its precision. Instead, radioactive beams used directly for treatment would yield an improved signal and a closer match with the dose fall-off, potentially enabling precise in vivo beam range monitoring. We have performed a treatment planning study to estimate the possible impact of the reduced range uncertainties, enabled by radioactive 11C ions treatments, on sparing critical organs in tumor proximity. Compared to 12C treatments, (i) annihilation maps for 11C ions can reflect sub- millimeter shifts in dose distributions in the patient, (ii) outcomes of treatment planning with 11C significantly improve and (iii) less severe toxicities for serial and parallel critical organs can be expected.
更多
查看译文
关键词
carbon reduction therapy,radioactive reduction beams
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要