Direct Quantitation of SARS-CoV-2 Using Droplet Digital PCR in Suspected Samples With Very Low Viral Load

crossref(2021)

引用 0|浏览3
暂无评分
摘要
Background: The reference test for SARS-CoV-2 detection is the reverse transcriptase real time PCR (real time RT-PCR). However, evidences reported that real time RT-PCR has a lower sensitivity compared with the droplet digital PCR (ddPCR) leading to possible false negative in low viral load cases. Methods: We used ddPCR for viral genes N1 and N2 on 20 negative (no detection) samples from symptomatic hospitalized COVID-patients presenting fluctuating real time RT-PCR results and 10 suspected samples (Ct value>35) from asymptomatic not hospitalized subjects. Results: ddPCR performed on RNA revealed 65% of positivity for at least one viral target in the hospitalized patients group of samples (35% for N1 and N2, 10% only for N1 and 20% only for N2) and 50% in the suspected cases (30% for N1 and N2, while 20% only for N2). On hospitalized patients’ samples, we applied also a direct ddPCR approach on the swab material, achieving an overall positivity of 83%. Conclusion: ddPCR, in particular the direct quantitation on swabs, shows a sensitivity advantage for the SARS-CoV-2 identification and may be useful to reduce the false negative diagnosis, especially for low viral load suspected samples.