Stress shielding effect after total hip arthroplasty varies between combinations of stem design and stiffness—a comparing biomechanical finite element analysis

International orthopaedics(2023)

引用 2|浏览2
暂无评分
摘要
Purpose Total hip arthroplasty (THA) has become a highly frequent orthopaedic procedure. Multiple approaches have been made to design the femoral component for THA with a mechanical behaviour as close as possible to a natural femur. The aim of this study was to compare different combinations of design and biomechanical properties of THA prostheses and their impact on stress shielding of the periprosthetic bone. Methods Virtual implantation of different stem designs (straight standard stem, straight short stem, anatomical short stem) by finite element analysis based on in vivo data from computer tomography was performed. For each stem, three grades of stiffness were generated, followed by a strain analysis. Results Reduction of stem stiffness led to less stress shielding. Implantation of an anatomical short-stem prosthesis with low stiffness provided the most physiological strain-loading effect ( p < 0.001). Conclusion A combination of a short and an anatomically designed stem with a low stiffness might provide a more physiological strain transfer during THA. Biomechanical properties of the femoral component for THA should be considered as a multifactorial function of dimensions, design, and stiffness.