A finite element study on the interactive effect between the damage of the cup-bone interface and the bone strain of hip implants under various fixation conditions.

Journal of the mechanical behavior of biomedical materials(2023)

引用 0|浏览0
暂无评分
摘要
Interfacial damage has a high impact on the loosening of the acetabular cup. However, monitoring this damage induced by the variations in loading conditions, such as the angle, amplitude, and frequency in vivo, is challenging. In this study, we evaluated the risk of loosening of the acetabular cup due to interfacial damages induced by the deviation in loading conditions and amplitudes. A three-dimensional model of the acetabular cup component was developed, and the interfacial crack growth between the cup and the bone was modeled using a fracture mechanics approach, which simulated the extent of interfacial damage and associated cup displacement. The interfacial delamination mechanism changed with the increasing inclination angle, wherein a fixation angle of 60° exhibited the largest area of contact loss. The compressive strain of embedding the simulated bone at the remaining bonding area accumulated as the lost contact area widened. Such interfacial damages, namely, the growth of the lost contact area and accumulated compressive strain in the simulated bone, promoted both embedding and rotational displacement of the acetabular cup. In the worst case of a fixation angle of 60°, the total displacement of the acetabular cup exceeded the limit of the modified safe zone, suggesting a quantitative risk of dislocation of the acetabular cup induced by the accumulated interfacial damage. Furthermore, nonlinear regression analyses between the degree of displacement of the acetabular cup and the extent of the two types of interfacial damage demonstrated that the interactive effect of the fixation angle with the loading amplitude showed a significant effect on increasing cup displacement. These findings suggest that proper control of the fixation angle during operation is useful in preventing the loosening of the hip joint.
更多
查看译文
关键词
Artificial hip joint, FEA, Interfacial fracture, Acetabular cup
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要