Intraoperative CBCT imaging in endovascular abdomen aneurysm repair - Optimization of exposure parameters using a stent phantom.

Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)(2023)

引用 0|浏览4
暂无评分
摘要
Cone beam computed tomography (CBCT) may provide essential additional image guidance to endovascular abdominal aneurysm repair (EVAR) operations but also significant radiation exposure to patients if scans are not carefully optimized. The purpose of our study was to define the image quality requirements for intraoperative EVAR CBCT imaging and to optimize the CBCT exposure parameters accordingly. A Multi-Energy CT phantom simulating a large patient was used by replacing the central phantom cylinder with a custom water-filled insert including an EVAR stent. Different exposure parameters covering a range of radiation qualities and dose levels were used to define the optimal image quality level regarding stent graft evaluation (compressed, bent, or collapsed). The radiation dose was measured with a calibrated air kerma-area product (KAP) meter and organ doses were calculated based on Monte Carlo simulations and a mathematical patient model. Based on the results, updated exposure parameters with the highest mean energy and lowest dose level available were recommended. With the updated protocol, the radiation exposure could be significantly decreased. The KAP value decreased from 9720 μGy·m2 to 440 μGy·m2 and reference point air kerma from 351 mGy to 16 mGy (a reduction of 96%) and organ doses of the organs in the irradiated region decreased on an average 91%. The new protocol resulted in acceptable clinical image quality based on testing with clinical cases.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要