Biomechanical analysis of a novel screw system with a variable locking angle in mandible angle fractures

Medical & biological engineering & computing(2023)

引用 0|浏览7
暂无评分
摘要
Locking plates nowadays represent an important treatment in bone trauma and bone healing due to its strong biomechanical properties. The purpose of this study was to both computationally and experimentally validate a novel screw locking system by comparing it to another locking system from state-of-the-art and to apply it in an environment of a fractured mandible. FEA was used to test both systems prior to experimental tests. The systems were locked in the plate holes at 0°, 10°, 15°, and 20°. Cyclic bending tests and push-out tests were performed in order to determine the stiffness and push-out forces of both locking systems. Finally, newly designed locking system was implemented in mandibular angle fracture. Control locking system was biomechanically superior in push-out test, but with no greater significance. In contrast, the new locking system showed greater stiffness by 17.3% at the deflection angle of 20° in cyclic tests, with lower values for other deflection angles. Similar values were displayed in fractured mandible angle environment. Greater stiffness of the new locking system in cyclic loading tests, together with polyaxiallity of the new locking screw, could lead to easier application and improved biomechanical stability of the mandible angle fractures. Graphical abstract
更多
查看译文
关键词
Locking screw,Variable locking angle,Finite element analysis,Stiffness,Mandible angle fractures
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要