Accuracy of digital and conventional systems in locating occlusal contacts: A clinical study.

The Journal of prosthetic dentistry(2023)

引用 0|浏览0
暂无评分
摘要
STATEMENT OF PROBLEM:The accuracy of methods used for locating occlusal contacts throughout the entire clinical procedure has been poorly studied. PURPOSE:The purpose of this clinical study was to determine the reproducibility and criterion validity for different methods of locating occlusal contacts. MATERIAL AND METHODS:Thirty-two adults with natural dentitions participated in this cross-sectional test-retest study. In total, occlusal contacts at maximum intercuspation were recorded by using 15 methods: silicone transillumination with Occlufast Rock (40, 50, 100, and 200 µm) and Occlufast CAD (40 and 50 µm); virtual occlusion (100, 200, 300, and 400 µm); articulating film (12-, 40-, 100-, and 200-µm-thick); and T-Scan III. Images of the occlusal records were scaled and calibrated spatially, and the occlusal contacts of the right posterior mandibular teeth were delimited by using the FIJI software program. Reproducibility was expressed as 95% confidence intervals (95% CI) of the percentage of agreement in the location of the occlusal contacts between images from the test sessions against retest sessions using the same method. Criterion validity was expressed as 95% CI of the percentage of agreement in the location of the occlusal contacts between images from the test sessions against images from Occlufast Rock (criterion standard). RESULTS:Occlufast Rock achieved 85% to 95% agreement in the location of the occlusal contacts between the 2 sessions, whereas Occlufast CAD, 200-µm articulating film, and T-Scan offered 79% to 86%, 68% to 75%, and 65% to 75% agreement, respectively. The most valid method was Occlufast CAD (74% to 80%) followed by the 200-µm articulating film (57% to 63%), 400-µm virtual occlusion (53% to 62%), 100-µm articulating film (52% to 60%), and T-Scan (48% to 56%). CONCLUSIONS:Conventional methods, such as 100- and 200-µm articulating film and digital methods, including 400 µm virtual occlusion and T-Scan, offer sufficient accuracy in locating the occlusal contacts. However, strategies are needed to improve accuracy.
更多
查看译文
AI 理解论文
溯源树
样例
生成溯源树,研究论文发展脉络
Chat Paper
正在生成论文摘要